Radiation detector resolution over a continuous energy range
An ion approach is demonstrated to determine energy resolution in both semiconductor detectors and scintillators over a continuous energy range. For semiconductors, the energy resolution of a silicon detector was measured as a function of helium ion energy, and the values from extrapolation to high energies are in good agreement with the literature data from alpha measurements. For scintillators,
