Verification of a pencil beam based treatment planning system: output factors for open photon beams shaped with MLC or blocks
The accuracy of monitor unit calculations from a pencil beam based, three-dimensional treatment planning system (3D TPS) has been evaluated for open irregularly shaped photon fields. The dose per monitor unit was measured in water and in air for x-ray beam qualities from 6 to 15 MV. The fields were shaped either with a multileaf collimator (MLC) or with customized alloy blocks. Calculations from t